Electronic converter for flowmeters MT03F Instructions manual R-MI-MT03F Rev.: 0 English version

Page 2

10 3.2.5 Analog output wiring Terminal 8 mA (positive, active output) 9 mA 10 mA (negative, passive output) The analog output is galvanicall

Page 3

11 3.2.6 Relay output wiring Terminal Description Relay 13 Normally open Relay 1 14 Normally closed Relay 1 15 Common Relay 1 16 No

Page 4

12 3.2.7 Remote reset input wiring Terminal 19 Contact 20 Contact These terminals correspond to an input that starts a batching process. A pote

Page 5 - 3 INSTALLATION

13 (Down / Left) To enter the preset values for a batching. To switch between flow rate, totalizer and fluid velocity screens. To change to the digit

Page 6

14 To change the access password of the programming menu, select "Programming" in the main menu and then "Password". 5 MAIN MENU T

Page 7 - - Free potential contact

15 When the "Password" option is selected, a screen that indicates the password status for this menu appears. Selecting “enable”, the scree

Page 8

16 Once in the installation menu, the first screen allows to choose between the different options. 6 INSTALLATION PARAMETERS Power on the electronic c

Page 9

17 6.2 K factor In this screen the K factor from the pulse transmitter (COVOL, TM44 or other) must be entered. This value should be found on the label

Page 10 - Passive output

18 Pulse input is compatible with all flowmeters series COVOL, and with any instrument that has an pulse output of one of the following types: - De 0

Page 11

19 If 4mA is selected, pressing the key (Enter) toggles between 4, 8, 12, 16 and 20 mA, allowing to check the adjustment of the analog output. NOTE: T

Page 12

2 Thank you for choosing the MT03F converter from MT03 series of Tecfluid S.A. This instruction manual allows the installation, programming and operat

Page 13 - 4 OPERATION

20 With the keys (Down / Left) and (Up), select Programming, and then validate with the key (Enter). If the converter has the password option enabled,

Page 14 - 5 MAIN MENU

21 7.4 Flow rate decimals In this screen the number of decimals for the flow rate indication can be selected. To select the number of decimals it mus

Page 15

22 7.5.1 Cut off A cut off flow rate can be programmed, that means, the flow rate below which the flow rate indication will be zero. This can avoid re

Page 16 - 6 INSTALLATION PARAMETERS

23 7.6.1.1 Batching If batching option is selected, the relay will change its state at the start or end of a batching process. For detailed informatio

Page 17

24 7.6.1.3 Deactivation The relay will be always at rest. 7.6.2 Analog output Before making a current calibration, be sure that the ammeter used for t

Page 18 - - Potential free contact

25 The first three screens show the relay output status, if they are activated. If the relay is in alarm mode, an “A” will appear in normal or inverse

Page 19 - 7 PROGRAMMING PARAMETERS

26 12 BATCHING A batching process is performed using the following parameters: - Preset value. Is the amount to be dispensed. The MT03F can work with

Page 20

27 - Relays. If configured as batching they change its state when the accumulated total reaches the preset value associated with each relay. From the

Page 21

28 13 WORKING SCREEN When exiting the menu, the display shows the working screen. To scroll between the six operating screens, press the key (Down /

Page 22

29 16 ASSOCIATED SOFTWARE WINSMETER MT03 Most of the steps in the preceding paragraphs can be done by means of the device associated software Winsmet

Page 23

3 TABLE OF CONTENTS 1 INTRODUCTION ... 5 2 RECEPTION ...

Page 24

30 Execute the program WinsmeterMT03 following the sequence Start – Programs – Tecfluid S.A. - WinsmeterMT03. 16.2 Port connection In the "Port&q

Page 25

31 16.3 Access to installation and programming In order to change the data contained in the "Install" tab, you must enter a password. The d

Page 26 - 10 SOFTWARE VERSION

32 In the installation window the parameters that adapts the converter to a sensor can be configured. To transfer data to the MT03F converter, press t

Page 27

33 Changing the parameters of this screen you can program the different functions of the equipment. As in the previous paragraph, to program this data

Page 28 - Fk n = Fk

34 16.5 Graphic This screen can graphically display the flow rate and partial counter progress. It is intended to be most useful in batching processes

Page 29

35 16.6 Datalogger In this screen the different process variables of the converter can be registered in a file. The time between samples can be set, a

Page 30

36 The process takes about 90 seconds, after which the message “Device programmed” will appear. From this moment, the converter MT03F already has the

Page 31

37 17 TECHNICAL CHARACTERISTICS Power supply 90 ... 265 VAC 50, 60 Hz 18 ... 36 VDC Power consumption:  5 VA Pulse input From 0 to VCC, where

Page 32

38 Housing Material: Box: Noryl SE1, GFN2, 701 black (1) Front frame: Lexan 920 black (1) Ingress protection: Back side: IP30 Front side:

Page 33

39 19 TROUBLESHOOTING Problem Probable cause Solution The flow rate is 0, the accumulated total counter and the totalizer do not increase their v

Page 34 - 16.5 Graphic

4 7.6 Outputs ... 22 7.6.1 Relay 1 and relay 2 ...

Page 35

40 WARRANTY Tecfluid S.A. guarantees all the products for a period of 24 months from their sale, against all faulty materials, manufacturing or perfor

Page 36

5 1 INTRODUCTION MT03F converters are flow rate indicators, totalizers and controllers for batching processes. The electronic circuit is based on the

Page 37

6 3.2 Electrical connection The connection of the equipment is via plug-in connectors, polarized to avoid mistakes when plugged in. The connectors hav

Page 38 - Hole panel

7 Before starting the wiring of the instrument, check that the supply voltage available is the same as marked on the label of the converter. Termina

Page 39 - 19 TROUBLESHOOTING

8 Terminal 4 Turbine 5 Turbine 3.2.3 Pickup (turbine) input wiring Open collector 5V power supply 5V … 24V pulses

Page 40

9 This input is compatible with all turbine flowmeters series TM and those instruments that provide a magnetic pickup type output whose voltage is bet